Design and Safety in Power Supplies for Advanced Medical Devices

Regulations and Challenges in Medical Device Design

The increasing adoption of advanced medical devices has led to the design and use of a wide range of complex medical equipment, both in hospital settings and for home telecare. As these devices are used in a variety of applications and configurations, designers face a critical challenge early in the design process: How to select or design medical devices, including their power supplies, to ensure safe and appropriate use?

This article explains the risk guidelines applicable to medical devices used with patients, the protective measures required for different types of medical devices, the designated use environments for these devices, and the electromagnetic compatibility (EMC) considerations crucial to designing power supplies for medical devices in any environment. These medical device design requirements are primarily established by IEC 60601-1 and its collateral standard, IEC 60601-1-2.

Regulations for Medical Devices and Medical Grade Power Supplies

The most important standard governing the design of medical devices is IEC 60601-1. This standard defines the general requirements for the basic safety and essential performance of a medical device. Compliance with IEC 60601-1 is essential for the marketing of any medical device in the USA. and worldwide, as the European (EN) and Canadian (CSA) versions of the standard are identical to IEC 60601-1. For EMC compliance, a medical device must comply with IEC 60601-1-2, Electromagnetic Perturbations – Requirements and Testing, which is a collateral standard to IEC 60601-1.

Basic Safety Requirements for Medical Grade Medical Devices and Medical Grade Power Supplies According to IEC 60601-1

Any medical grade power supply must be designed or selected so that the basic safety and essential performance of medical devices are not compromised. IEC 60601-1 ensures patient safety by requiring means of protection, abbreviated as MOP. MOPs are the means to reduce the risk of electric shock that can be achieved through insulation, air gaps, creepage distances, impedances and protective earth connections.

Protective finishes such as varnishing, enameling, oxidation and similar coatings that can be re-plasticized at temperatures expected during operation (including sterilization) are not considered as means of protection (MOP). MOPs are further classified into MOOPs (means of operator protection) and MOPPs (means of patient protection). Design requirements, such as creepage distance and clearances, are different for MOOP and MOPP.

Compliance with MOPs is verified by the dielectric strength test for solid insulation as MOP. When creepage distance and clearance values are claimed as MOP, they must comply with the limits specified in IEC 60601-1. Where protective earth connections form a means of patient protection, they must comply with the grounding test requirements of the standard.

Choosing a Means of Protection (MOP)-Based Power Supply

The obvious question for a designer is how to choose a medical grade power supply based on means of protection (MOP) requirements. According to IEC 60601-1, any medical device, and therefore its power supply, must provide two means of medical protection to prevent applied and other accessible parts from posing a risk.

Medical grade power supplies on the market comply with 2 x MOPP input to output and 1 x MOPP input to ground. However, most power supplies employ a grounded output operating insulation that typically only achieves a 500 VAC/VDC rating. This is insufficient to meet the 1500 VAC 1 x MOP insulation voltage test, even before considering creepage distances and clearances, and excessive leakage current from output to ground resulting from an input to output insulation capacitance that is too high. In these cases, a secondary DC/DC converter is required to provide additional MOPP.

Role of Leakage Current in Selecting a Power Supply

Leakage current is defined in IEC 60601-1 as current that is not functional. The following leakage currents are classified as earth leakage current, contact current and patient leakage current.

Earth leakage current is the current flowing through the protective earth conductor, but does not represent a risk to the patient or operator, as both are protected by specifying appropriately low values for patient leakage current and touch current under normal conditions and under relevant fault conditions, including interruption of the protective earth conductor. However, excessive leakage current to ground could represent a potential problem for the facility’s grounding system and any circuit breakers operated by current unbalance detectors.

Contact current is the leakage current flowing from the enclosure or parts of the enclosure, excluding patient connections, accessible to any operator or patient in normal use, through an external path other than the protective earth conductor, to ground or to another part of the enclosure.

Patient leakage current is the current flowing from the patient connections through the patient to ground, or the current that originates from the unintentional occurrence of a voltage from an external source on the patient and flows from the patient through the patient connections of an applied f-type part to ground.

An important term to understand is the concept of the Applied Party, which is part of the medical team that comes into physical contact with the patient to perform its function. For example, one of the applied parts of an ECG machine is the ECG electrode. The applied part of a medical device can be of type B (body), BF (floating body) or CF (floating heart).

Type B applied provides some protection against electric shock but is the least stringent in terms of allowable patient leakage current and patient auxiliary current. Type B applied parts are not suitable for direct-to-heart applications. Type BF are electrically floating applied parts that meet the requirements specified in this standard to provide a higher degree of protection against electric shock than Type B applied parts. Type CF are electrically floating applied parts that meet the requirements specified in this standard to provide a higher degree of protection against electric shock than type BF applied parts.

Sometimes standard power supplies, including those that are medically approved, do not meet the requirements for grounded outlet isolation or patient leakage current. An additional isolation transformer at the power supply input or a secondary DC/DC converter may be required to ensure compliance.

Conclusion

There are many power supplies that are listed as medical grade only because they provide two means of patient protection, but may not meet the additional requirements of IEC 60601-1. Designing or selecting a power supply for medical devices, especially those involving electrical contact with the patient, requires addressing a dual challenge: minimizing leakage currents under normal conditions and providing ground isolation to protect the patient in the event of a fault.

Achieving lower leakage currents, especially for cardiac (BF-qualified) applications, presents its own set of challenges. In addition to the need to significantly increase the spacing between internal components on the secondary side to meet high insulation voltage requirements, there is also a conflict between achieving low leakage currents and ensuring low emissions.

It is also crucial to minimize both differential noise and common-mode noise by using a low-noise topology, as well as to reduce line frequency ripple in the primary circuits. The designer should not aim for a lower leakage current rating than necessary, as this could result in higher emissions. After integrating the power supply into the system, leakage current must be measured for the overall system, but having a medical grade power supply gives the designer confidence in achieving compliance with IEC 60601-1.